Invacare 600 lb. Bariatric Bed, Head motor extension cable used on component numbers BAR5490 and BAR5490IVC. These components are part of the following Bed Model Number(s): BAR600, BARPKGCA, BAR600IVC, BARPKGIVC-1633, and BARPKG.
Recall
- Recall Number
- Z-0187-2012
- Event Number
- 59461
- Firm
- Invacare Corporation
- FEI Number
- 1525712
- Product Code
- LLI
- Status
- Terminated
- Root Cause
- Labeling design
- Initiated
- June 23, 2011
- Posted
- November 14, 2011
- Terminated
- July 2, 2013
- Address
- 1200 Taylor St, Elyria, OH, 44035-6248
Description
Invacare 600 lb. Bariatric Bed, Head motor extension cable used on component numbers BAR5490 and BAR5490IVC. These components are part of the following Bed Model Number(s): BAR600, BARPKGCA, BAR600IVC, BARPKGIVC-1633, and BARPKG.
Invacare Corporation decided to recall the products because risk analysis reveals that miskeying or disconnection of the head motor extension cable during bed set-up may result in product damage, or in rare instances, the potential for fire.
On 9/19/11, Invacare sent Recall Notificaiton Letters to their customers. They instructed their customers to quarantine the product, complete and return a response form.
Distribution Nationwide and to United Arab Emirates, Dominican Republic, Egypt, Guam, Mexico, & Virgin Islands.
35,747