LIFEPAK(R) defibrillation electrodes, Catalog # 11101-000016, MIN #3202380-006 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.
Recall
- Recall Number
- Z-0186-2018
- Event Number
- 78303
- Firm
- Physio-Control, Inc.
- FEI Number
- 3015876
- Product Code
- MKJ
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- October 27, 2017
- Posted
- December 12, 2017
- Terminated
- June 12, 2019
- Address
- 11811 Willows Rd NE, Redmond, WA, 98052-2003
Description
LIFEPAK(R) defibrillation electrodes, Catalog # 11101-000016, MIN #3202380-006 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.
The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
The firm Physio Control, initiated their field action on 10/27/2017 by issuing a press release. The firm followed this with a recall notice dated 10/23/2017, to the direct consignees on 11/03/2017. The letter described the product, explained the problem and provided the following instructions: Quarantine the affected lots;contact Physio-Control at 1-866-231-1220, 6:00 A.M. to 4:00 P.M. (Pacific), Monday Friday to arrange for replacement of your unused electrodes and "Upon receipt of this notification, post a copy of the enclosed Correct Electrode and Packaging Labeling with each of your AEDs, which shows the correct placement of the electrodes." Physio_Control will provide replacement products for all unused affected Electrodes. Should you have any questions about this Product Recall, please contact us at 1-866-231-1220, 6:00 A.M. to 4:00 P.M. (Pacific), Monday Friday.
Worldwide distribution: US (nationwide) including Puerto Rico and Guam and countries of: Canada, Austria, Belgium, Czech Republic; Denmark, Finland, France, French Polynesia, Germany, Hungary, Ireland, Italy, Latvia, Netherlands, Australia, China, Hong Kong, Japan, Argentina, Mexico, Norway, Panama, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, and New Zealand.
14,219 units in total