FDA Recall Terminated

Combiset Blood Line with BVM for Hemodialysis use Catalog Number: 03-2965-9. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.

Recall: Z-0185-2014 · Initiated August 29, 2013

Recall

Recall Number
Z-0185-2014
Event Number
65739
Firm
Fresenius Medical Care Holdings, Inc.
FEI Number
3001451489
Product Code
ONW
Status
Terminated
Root Cause
Labeling design
Initiated
August 29, 2013
Posted
November 8, 2013
Terminated
February 11, 2021
Address
920 Winter St, Waltham, MA, 02451-1521

Description

Combiset Blood Line with BVM for Hemodialysis use Catalog Number: 03-2965-9. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.

Reason

Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss

Action

Fresenius issued a Field Notification Letter dated 8/29/13 to Patient Clinics/ associated at home patients and provides additional detail instructions to ensure the connection to the transducer protector bloodline female luer interface is secure. This field advisory notice sent by Certified Mail, Return Receipt Requested. Accounts requested to return both the receipt and the completed reply form (enclosed) to confirm receipt of the notice. If you have any questions regarding this field advisory, please contact FMCNA Customer Service Team at 1-800-323-5188. Customer Service will be available Monday-Friday 7am-6pm CST.

Distribution

Worldwide Distribution-USA (nationwide) and the country of Canada.

Quantity

24,202 cases