FDA Recall Terminated

Encore Reverse Shoulder Prosthesis (RSP) System, Standard Size 40 RSP Humeral Socket Insert; Catalog #: 508-00-040; Sterile; Manufactured by Encore, Medical, L.P., Austin, TX 78758.

Recall: Z-0185-2008 · Initiated August 8, 2007

Recall

Recall Number
Z-0185-2008
Event Number
44800
FEI Number
1000116912
Product Code
KWS
Status
Terminated
Root Cause
Other
Initiated
August 8, 2007
Posted
December 19, 2007
Terminated
March 10, 2008
Address
9800 Metric Blvd, Encore Medical, Lp, Austin, TX, 78758

Description

Encore Reverse Shoulder Prosthesis (RSP) System, Standard Size 40 RSP Humeral Socket Insert; Catalog #: 508-00-040; Sterile; Manufactured by Encore, Medical, L.P., Austin, TX 78758.

Reason

Devices Mislabeled; one lot of Size 36 Standard RSP Socket Inserts were labeled as Size 40 Standard RSP Sockets and vice versa.

Action

Encore notified their sales reps (direct accounts) via telephone and overnight letter on 08/08/07. Encore sales reps were asked to return all outstanding devices, for which they would be given replacements. Notification of implanting surgeons was done by phone and letter (Urgent : Device Recall) between 08/07 and 09/12/07. Implanting surgeons notified of recall and asked to closely monitor patients who had received and already been implanted with the affected lots. If explant was required decision of the surgeons, they were asked to notify the recalling firm.

Distribution

Nationwide:Direct accounts and surgeons in the following states: CA, GA, IA, MA, MD, MI, MO, SC, TX and UT.

Quantity

25 units.