Trumpf Universal Imaging Extension (OEM) Catalog NUmber: F-PMT-A2
Recall: Z-0184-04
·
Initiated November 6, 2003
Recall
- Recall Number
- Z-0184-04
- Event Number
- 27655
- Firm
- The OR Group
- FEI Number
- 1221538
- Product Code
- FPS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 6, 2003
- Posted
- December 10, 2003
- Terminated
- July 25, 2005
- Address
- 531 Main St, Acton, MA, 01720-3934
Description
Trumpf Universal Imaging Extension (OEM) Catalog NUmber: F-PMT-A2
Reason
Imaging table with extension may fail if the table leg is incorrectly positioned and the load exceeds the maximum weight of 400 lb
Action
The OR Group forwarded by 2nd day mail on 11/6/03 the repair kit and instructions to the accounts.
Distribution
Nationwide. CA, FL, GA, IA MI, MT, NH, OH, NJ, NY, SC, UT, TX, VA
Quantity
1 unit