FDA Recall Terminated

GE OEC, Insta Trak 3500 Plus with the Cranial Application for Image Guided Surgery, GE Healthcare, Surgery, Salt Lake City, UT 84116

Recall: Z-0183-2008 · Initiated October 18, 2007

Recall

Recall Number
Z-0183-2008
Event Number
45441
Firm
GE OEC Medical Systems, Inc
FEI Number
1720753
Product Code
LLZ
Status
Terminated
Root Cause
Device Design
Initiated
October 18, 2007
Posted
January 23, 2008
Terminated
August 10, 2012
Address
384 Wright Brothers Dr, Salt Lake City, UT, 84116-2862

Description

GE OEC, Insta Trak 3500 Plus with the Cranial Application for Image Guided Surgery, GE Healthcare, Surgery, Salt Lake City, UT 84116

Reason

Crack, Fracture or Detachment: The Transmitter Arm for the Mayfield-compatible Transmitter may become detached from the unit and enter into the surgical field.

Action

Consignees were sent an Urgent Recall Notice via mail dated 10/11/07, notifying the customer to examine the plastic end of the transmitter arm and to not use the transmitter arm, if any cracks or damage are observed. Also to stop using Steam Sterilization method and to utilize the other Sterilization methods indicated in the Operators Manual. The root cause is still under investigation GEHC Surgery will provide, at no charge, a solution to this issue when it becomes available.

Distribution

Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IL, MA, MD, MI, MS, NC, NY, OK, OR, PA, SC, TX, UT, VA, WI, and WV, and countries of Canada, China, France, Hong Kong, Japan, Mexico, Spain, and Switzerland.

Quantity

169 units