M3001A Philips Multi Measurement Server (MMS)
Recall
- Recall Number
- Z-0183-2007
- Event Number
- 36489
- Firm
- Philips Medical Systems
- FEI Number
- 1218950
- Product Code
- DQA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 21, 2006
- Posted
- November 16, 2006
- Terminated
- August 11, 2020
- Address
- 3000 Minuteman Road, Andover, MA, 01810-1032
Description
M3001A Philips Multi Measurement Server (MMS)
Unexpected pulse oximetry (Sp02) readings (100%) over time when a sensor is not attached to the patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM
Philips issued a Customer Letter and Urgent Device Correction letter to consignees on 09/21/06. Users were notified of the problem and provided with Instructions to Mitigate Risk. A software upgrade to: version D.00.61 and Sp02 FW A.01.46 is mandatory.
Worldwide, including USA, Canada, Australia, China, Hong Kong, India, Indonesia, Japan, Phillipines, Singapore, Taiwan, Thailand, Vietnam, Argentina, Brazil, Chile, Colombia, El Salvador, Jamaica, Mexico, Nicaragua. Panama, Paraguay, Peru, Venezuela, Austria, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Germany, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Lithuania, Luxembourg, Macedonia, Mauritania, Mauritius, Netherlands, Norway, Oman, Slovenia, South Africa, Spain, Switzerland, Turkey, UAE, UK, Yemen, Zambia, Zimbabwe
69696 units