FDA Recall Terminated

M3001A Philips Multi Measurement Server (MMS)

Recall: Z-0183-2007 · Initiated September 21, 2006

Recall

Recall Number
Z-0183-2007
Event Number
36489
Firm
Philips Medical Systems
FEI Number
1218950
Product Code
DQA
Status
Terminated
Root Cause
Other
Initiated
September 21, 2006
Posted
November 16, 2006
Terminated
August 11, 2020
Address
3000 Minuteman Road, Andover, MA, 01810-1032

Description

M3001A Philips Multi Measurement Server (MMS)

Reason

Unexpected pulse oximetry (Sp02) readings (100%) over time when a sensor is not attached to the patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM

Action

Philips issued a Customer Letter and Urgent Device Correction letter to consignees on 09/21/06. Users were notified of the problem and provided with Instructions to Mitigate Risk. A software upgrade to: version D.00.61 and Sp02 FW A.01.46 is mandatory.

Distribution

Worldwide, including USA, Canada, Australia, China, Hong Kong, India, Indonesia, Japan, Phillipines, Singapore, Taiwan, Thailand, Vietnam, Argentina, Brazil, Chile, Colombia, El Salvador, Jamaica, Mexico, Nicaragua. Panama, Paraguay, Peru, Venezuela, Austria, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Germany, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Lithuania, Luxembourg, Macedonia, Mauritania, Mauritius, Netherlands, Norway, Oman, Slovenia, South Africa, Spain, Switzerland, Turkey, UAE, UK, Yemen, Zambia, Zimbabwe

Quantity

69696 units