FDA Recall
Terminated
Ronci Surgical Universal Imaging Extension used with the Ronci Pain Management Table System Catalog Number: R-934-25A
Recall: Z-0182-04
·
Initiated November 6, 2003
Recall
- Recall Number
- Z-0182-04
- Event Number
- 27655
- Firm
- The OR Group
- FEI Number
- 1221538
- Product Code
- FPS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 6, 2003
- Posted
- December 10, 2003
- Terminated
- July 25, 2005
- Address
- 531 Main St, Acton, MA, 01720-3934
Description
Ronci Surgical Universal Imaging Extension used with the Ronci Pain Management Table System Catalog Number: R-934-25A
Reason
Imaging table with extension may fail if the table leg is incorrectly positioned and the load exceeds the maximum weight of 400 lb
Action
The OR Group forwarded by 2nd day mail on 11/6/03 the repair kit and instructions to the accounts.
Distribution
Nationwide. CA, FL, GA, IA MI, MT, NH, OH, NJ, NY, SC, UT, TX, VA
Quantity
17 units