Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
Recall
- Recall Number
- Z-0180-2026
- Event Number
- 97659
- Firm
- American Contract Systems Inc.
- FEI Number
- 3007069947
- Product Code
- OEQ
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- September 4, 2025
- Posted
- October 16, 2025
- Address
- 1601 Se Gateway Dr, Ste 120, Grimes, IA, 50111-5223
Description
Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
32,433 total