FDA Recall Open, Classified

Entellus Medical REF SSA-102 SN ******* Shaver System Foot Pedal - Product Usage: The Entellus Medical Shaver System has been designed for shaping bone and for the resection of soft and hard tissues as part of surgical procedures in the areas of otorhinolaryngology.

Recall: Z-0180-2021 · Initiated September 11, 2020

Recall

Recall Number
Z-0180-2021
Event Number
86419
Firm
Bien-Air Surgery Sa rue de l' ouest 2b Le Noirmont Switzerland
FEI Number
3003800209
Product Code
ERL
Status
Open, Classified
Root Cause
Process control
Initiated
September 11, 2020

Description

Entellus Medical REF SSA-102 SN ******* Shaver System Foot Pedal - Product Usage: The Entellus Medical Shaver System has been designed for shaping bone and for the resection of soft and hard tissues as part of surgical procedures in the areas of otorhinolaryngology.

Reason

The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.

Action

On 09/14/2020, the firm send out an "URGENT: Medical Device Correction" notification informing customer of the potential for the magnet located inside the foot pedal displaces or comes off during a surgery and therefore may compromise the device's safety. The use of the incriminated devices may entail a risk of unattended start or unstoppable motor when acting on the foot pedal. The Recalling Firm is informing customers of a curative and a corrective action to prevent the magnet from coming off: -Curative Action: 1) Inspect your inventory for the product numbers/serial numbers above. 2) Quarantine any of the affected products until performing the following actions 3) Open the foot pedal rocker to have access to the incriminated magnet by pulling the black part 4) Carefully inspect the position of the magnet using illustrations provided in the Customer Notification If the magnet is partially or full off, Customers are instructed to keep the foot pedals in quarantine (Do Not Use It Anymore), and contact the Recalling Firm's repair center and send the foot pedal back for repair. 5) If the magnet is well position, then carefully clean the plastic area around the magnet with a disinfectant wipe (e.g. alcoholic solution) and a swab to correctly clean the corners 6) Make sure the surface is clean and dry before applying an adhesive tape (e.g. Scotch, width between 10-15mm, length between 55-60mm) as described in the illustrations provide in the Customer Notification THE FOLLOWING ACTIONS MUST BE PERFORMED BEFORE EACH SURGERY 7) Carefully inspect that the adhesive tape is correctly holding the magnet. If needed, remove the tape and restart the process from step 4. 8) Close the foot pedal rocker, connect the foot pedal cable to the control unit and switch it ON 9) Proceed with the following tests before performing the surgery a. Select one motor (BASCH, 80K, NANO, RAPIDO) or one handpiece (S120, OSSEOSTAP, PERFO) depending on your needs b. Hold the motor

Distribution

US Nationwide distribution including in the states of CA, MI and MN.