FDA Recall Terminated

Reverse Shoulder Prosthesis Stem Impaction Fixture REF 804-03-053 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.

Recall: Z-0180-2016 · Initiated September 14, 2015

Recall

Recall Number
Z-0180-2016
Event Number
72229
FEI Number
1000116912
Product Code
HWA
Status
Terminated
Root Cause
Device Design
Initiated
September 14, 2015
Terminated
November 4, 2016
Address
9800 Metric Blvd, Encore Medical, Lp, Austin, TX, 78758-5445

Description

Reverse Shoulder Prosthesis Stem Impaction Fixture REF 804-03-053 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.

Reason

During the assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.

Action

DJO Global sent an Urgent Field Safety Notice dated September 15, 2015, to all affected consignees. The notice identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed become familiar with the new techniques and notify all of their applicable users of the RSP Monoblock and Turon stems of these revised methods. When each user has been notified of these new techniques, they should complete the form that was provided with the notification indicating that they have been informed of these changes and understand the new methods. Consignees were asked to complete and return forms. Consignees with questions should call 512-834-6255.

Distribution

Nationwide Distribution

Quantity

634 units