FDA Recall Terminated

ConMed Linvatec Sternum Saw Collet Nut, Catalog Number 5059-09: Used on ConMed Linvatec Sternum Saw Handpieces (ConMed Linvatec, Hall. Versipower, Versipower Plus Handpieces) and sold individually.

Recall: Z-0179-2007 · Initiated December 20, 2005

Recall

Recall Number
Z-0179-2007
Event Number
34438
Firm
Linvatec Corp.
FEI Number
1017294
Product Code
KFK
Status
Terminated
Root Cause
Other
Initiated
December 20, 2005
Posted
November 16, 2006
Terminated
July 11, 2012
Address
11311 Concept Blvd, Largo, FL, 33773-4908

Description

ConMed Linvatec Sternum Saw Collet Nut, Catalog Number 5059-09: Used on ConMed Linvatec Sternum Saw Handpieces (ConMed Linvatec, Hall. Versipower, Versipower Plus Handpieces) and sold individually.

Reason

The Sternum Saw Collet Nut that is used to insert and hold the saw blade in place may not always allow for the easy insertion of a saw blade into the handpiece for use. The nut needs to be in a certain position in order for the blade to be inserted and this may delay the operation of the device especially in emergency situations.

Action

Linvatec will send each consignee a recall package that will include: a 12/20/2005 recall letter and reply form, A product correction instruction sheet , a new Sternum Saw Collet Nut and an assembly manual. Linvatec will also contact each consignee by phone or email. All consignees that have not responded by 30 days after the first mailing will be re-contacted by the firm by mail, phone or email.

Distribution

Worldwide, including USA, Mexico, and Canada.

Quantity

11424 units