FDA Recall Terminated

FHC microTargeting Power Assist System used with the micro Targeting Drive System, models: 66-DA-ME; 66EL-MC; 66EL-MS;66EL-MS-02; 66-EL-MS-03; 66-EL-RM; MT-DS-AM; MT-DS-MD.

Recall: Z-0179-06 · Initiated October 31, 2005

Recall

Recall Number
Z-0179-06
Event Number
33946
Firm
FHC, Inc.
FEI Number
3002250546
Product Code
HAW
Status
Terminated
Root Cause
Other
Initiated
October 31, 2005
Posted
November 23, 2005
Terminated
June 7, 2006
Address
9 Main St, Bowdoinham, ME, 04008-4418

Description

FHC microTargeting Power Assist System used with the micro Targeting Drive System, models: 66-DA-ME; 66EL-MC; 66EL-MS;66EL-MS-02; 66-EL-MS-03; 66-EL-RM; MT-DS-AM; MT-DS-MD.

Reason

Power Assist microTargeting System may advance beyond intended target

Action

FHC Inc. and Medtronic notified customers by letter and instructs the customer to return their Motor Option for upgrade to include the following components: 1) Handheld Remote Control, 2) Motor Controller Unit with Display, 3) Motor, and 4) Medical Power Adapter. (See pictures at the end of this communication). Until the Motor Option has been updated, instructions to customers are: Prior to each clinical use: Ensure the microTargeting Drive system is calibrated before each clinical use per the User Guide. Check the microTargeting Drive system functionality over a range of at least 10mm movement. Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. During clinical use: Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. Every 100 uses or annually: Have your microTargeting Drive system checked by the manufacturer per the User Guide. Have your equipment support department check or tighten the MicroTargeting Drive Motor Option Remote Control Knob set screws, tighten as needed, and then re-calibrate.

Distribution

Nationwide Foreign: South Africa, Taiwan, Hong Kong, China, France

Quantity

70 units