Bien-Air Surgery REF 1600686-001 SN ******* OSSEOSTAP Footcontrol - Product Usage: an active medical device that convert electrical energy into mechanical rotation to a but during Ear, Nose and Throat (ENT) surgery.
Recall
- Recall Number
- Z-0178-2021
- Event Number
- 86419
- Firm
- Bien-Air Surgery Sa rue de l' ouest 2b Le Noirmont Switzerland
- FEI Number
- 3003800209
- Product Code
- ERL
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- September 11, 2020
Description
Bien-Air Surgery REF 1600686-001 SN ******* OSSEOSTAP Footcontrol - Product Usage: an active medical device that convert electrical energy into mechanical rotation to a but during Ear, Nose and Throat (ENT) surgery.
The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.
On 09/14/2020, the firm send out an "URGENT: Medical Device Correction" notification informing customer of the potential for the magnet located inside the foot pedal displaces or comes off during a surgery and therefore may compromise the device's safety. The use of the incriminated devices may entail a risk of unattended start or unstoppable motor when acting on the foot pedal. The Recalling Firm is informing customers of a curative and a corrective action to prevent the magnet from coming off: -Curative Action: 1) Inspect your inventory for the product numbers/serial numbers above. 2) Quarantine any of the affected products until performing the following actions 3) Open the foot pedal rocker to have access to the incriminated magnet by pulling the black part 4) Carefully inspect the position of the magnet using illustrations provided in the Customer Notification If the magnet is partially or full off, Customers are instructed to keep the foot pedals in quarantine (Do Not Use It Anymore), and contact the Recalling Firm's repair center and send the foot pedal back for repair. 5) If the magnet is well position, then carefully clean the plastic area around the magnet with a disinfectant wipe (e.g. alcoholic solution) and a swab to correctly clean the corners 6) Make sure the surface is clean and dry before applying an adhesive tape (e.g. Scotch, width between 10-15mm, length between 55-60mm) as described in the illustrations provide in the Customer Notification THE FOLLOWING ACTIONS MUST BE PERFORMED BEFORE EACH SURGERY 7) Carefully inspect that the adhesive tape is correctly holding the magnet. If needed, remove the tape and restart the process from step 4. 8) Close the foot pedal rocker, connect the foot pedal cable to the control unit and switch it ON 9) Proceed with the following tests before performing the surgery a. Select one motor (BASCH, 80K, NANO, RAPIDO) or one handpiece (S120, OSSEOSTAP, PERFO) depending on your needs b. Hold the motor
US Nationwide distribution including in the states of CA, MI and MN.
Total US = 1120 foot pedals