FDA Recall Terminated

Biograph mCT: a) S(64)-3R, Material Number 10248669 b) S (20) -3R, Material Number 10507786 c) Flow 20-4R, Material Number 10528958 d) m Flow 64-4R, Material Number 10529161 e) mCT S(40)-4R, Material Number 10248671 f) mCT S(64)-4R, Material Number 0248672 g) mCT S(20)-3R, Material Number 10534159 h) mCT Flow Edge-4R, Material Number 10528955 i) mCT Flow 40-3R, Material Number 10529158 j) mCT Flow 64-3R, Material Number 10529160 k) mCT X-4R, Material Number 10248670 l) mCT X-3R, Material Number 10248673 m) mCT S(20)-4R, Material Number 10534160 n) mCT S(40)-3R, Material Number 10248668 o) mCT Flow 40-4R, Material Number 10529159 p) mCT Flow 20-3R, Material Number 10528956 q) mCT Flow Edge-3R, Material Number 10528954 The Siemens Biograph Horizon and Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Recall: Z-0178-2019 · Initiated September 28, 2018

Recall

Recall Number
Z-0178-2019
Event Number
81144
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
1000514697
Product Code
KPS
Status
Terminated
Root Cause
Process control
Initiated
September 28, 2018
Terminated
February 11, 2021
Address
2501 Barrington Rd, Hoffman Estates, IL, 60192-2061

Description

Biograph mCT: a) S(64)-3R, Material Number 10248669 b) S (20) -3R, Material Number 10507786 c) Flow 20-4R, Material Number 10528958 d) m Flow 64-4R, Material Number 10529161 e) mCT S(40)-4R, Material Number 10248671 f) mCT S(64)-4R, Material Number 0248672 g) mCT S(20)-3R, Material Number 10534159 h) mCT Flow Edge-4R, Material Number 10528955 i) mCT Flow 40-3R, Material Number 10529158 j) mCT Flow 64-3R, Material Number 10529160 k) mCT X-4R, Material Number 10248670 l) mCT X-3R, Material Number 10248673 m) mCT S(20)-4R, Material Number 10534160 n) mCT S(40)-3R, Material Number 10248668 o) mCT Flow 40-4R, Material Number 10529159 p) mCT Flow 20-3R, Material Number 10528956 q) mCT Flow Edge-3R, Material Number 10528954 The Siemens Biograph Horizon and Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Reason

The terminals on the plug may become loose and overheat.

Action

The firm, Siemens Healthineers, disseminated the " Customer Safety Advisory Notice" on 09/28/2018 to its customers. The notice described the product, problem and actions to be taken. The notice stated the potential problem and how to identify it. The customers were instructed to do the following: in the event overheating occurs or notice a distinct burning odor; stop use of your system and contact Siemens Healthineers service representatives.The issue will be addressed through a service visit to check if the terminal screws are torqued to the proper specification. If you have any questions, please contact your local Siemens Healthineers service representatives at the following numbers: America: 1-800-888-7436 Europe, Middle East, and Africa: +49 9131 940 4000 Asia and Australia: +86(21)3811 2121

Distribution

Worldwide Distribution: US (nationwide) and countries of: Argentina, Austria, Australia, Azerbaijan, Belgium, Bulgaria, Brunnel, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Dominican Rep, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Republic Korea, Lebanon, Sri Lanka, Macedonia, Mexico, Malaysia, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Paraguay, Qatar, Romania, Russian Fed, Saudi Arabia, Singapore, Slovenia, Slovakia, Spain, South Africa, Switzerland, Thailand, Turkey, Taiwan, UAE, United Kingdom and Vietnam.

Quantity

890 units