FDA Recall Terminated

RESPIRONICS V60 VENTILATOR

Recall: Z-0176-2010 · Initiated October 26, 2009

Recall

Recall Number
Z-0176-2010
Event Number
53676
Firm
Respironics California Inc
FEI Number
2518422
Product Code
MNT
Status
Terminated
Root Cause
Device Design
Initiated
October 26, 2009
Posted
November 11, 2009
Terminated
October 13, 2011
Address
2271 Cosmos Ct, Carlsbad, CA, 92011-1517

Description

RESPIRONICS V60 VENTILATOR

Reason

The recall was initiated to correct an issue with a component on the V60 Ventilator Power Management (PM) Printed Circuit Board (PCB). A component on the PCB has the potential to break off or become damaged if the PCB is removed from the ventilator. If this occurs, the ventilator may not properly switch between back up battery power and AC power. This can result in power loss during ventilation fo

Action

The recall notification was initiated October 26, 2009 with the firm mailing a Urgent Medical Device Field Correction letter to the affected customers. The letter informed the customers of the reason for the notice and how Respironics Field Service Technicians were going to correct this issue. The letter also gave the customers instructions to follow pending the completion of the field inspection/replacement. For questions or additional information contact Respironics US Customer Service at 1.800.345.6443, Option #5 then Option #1.

Distribution

Nationwide and Internationally to : Libya, Hong Kong, France, Italy, India, Spain, United Kingdom, Germany, Netherlands and Saudi Arabia

Quantity

524