FDA Recall Terminated

Stryker brand Neptune waste management system, Neptune Rover, 15 amp plug; model 0700-003-000.

Recall: Z-0175-05 · Initiated October 12, 2004

Recall

Recall Number
Z-0175-05
Event Number
30208
Firm
Stryker Instruments Div. of Stryker Corporation
FEI Number
1811755
Product Code
FYD
Status
Terminated
Root Cause
Other
Initiated
October 12, 2004
Posted
November 4, 2004
Terminated
March 28, 2005
Address
4100 E. Milham, Kalamazoo, MI, 49001

Description

Stryker brand Neptune waste management system, Neptune Rover, 15 amp plug; model 0700-003-000.

Reason

The grounding pin may break inside of the electrical plug, resulting in a shock hazard.

Action

Hospital customers will be sent a recall notice dated 11/09/04, new plugs for installation on their units, and instructions for installation.

Distribution

United States and Canada.

Quantity

365