FDA Recall Terminated

Asahi AM-BIO D Series Hollow Fiber Dialyzers; a dry model single or multiple use dialyzer consisting of hollow fiber membranes of alkyl ether polymer grafted cellulose housed within a plastic casing of styrene butadiene block polymer, sterilized by gamma radiation before shipment; Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Made in Japan; 12 units per case; Model AM-BIO-75D - 1.5 m2 surface area and Model AM-BIO-100D - 2.0 m2 surface area

Recall: Z-0174-2007 · Initiated October 12, 2006

Recall

Recall Number
Z-0174-2007
Event Number
36430
Firm
Asahi Medical Co., Ltd
FEI Number
1000599821
Product Code
MSE
Status
Terminated
Root Cause
Other
Initiated
October 12, 2006
Posted
November 14, 2006
Terminated
November 28, 2007
Address
3100 Dundee Rd, Ste 201/202, Northbrook, IL, 60062-2437

Description

Asahi AM-BIO D Series Hollow Fiber Dialyzers; a dry model single or multiple use dialyzer consisting of hollow fiber membranes of alkyl ether polymer grafted cellulose housed within a plastic casing of styrene butadiene block polymer, sterilized by gamma radiation before shipment; Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Made in Japan; 12 units per case; Model AM-BIO-75D - 1.5 m2 surface area and Model AM-BIO-100D - 2.0 m2 surface area

Reason

There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle.

Action

Asahi sent Safety Alert letters on 10/12/06 to their re-use dialyzer customers advising them of the reports of difficulty separating the blood set from the dialyzer and not to use the Gambro Cartridge Blood Set with Asahi Reuse/Multiple Use Dialyzers in a reuse situation. Enclosed were copies of the temporary revision Instructions For Use with the following warning added at the top of the page: 'Additional Warning for Reuse/Multiple Use Dialyzers: Ashai Reuse/Multiple use dialyzers must not be used in connection with the Gambro cartridge Blood Set in a reuse environment.' Any questions were directed to Karen Borling, RN at 847-498-8500.

Distribution

Nationwide

Quantity

100,000 for all models