FDA Recall Open, Classified

BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BD FACSCanto II, and BD FACSCanto flow cytometers to determine the percentages and absolute counts of human lymphocyte subsets in peripheral whole blood for immunophenotyping.

Recall: Z-0173-2023 · Initiated October 3, 2022

Recall

Recall Number
Z-0173-2023
Event Number
90866
Firm
Becton, Dickinson and Company, BD Biosciences
FEI Number
2916837
Product Code
GKZ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
October 3, 2022
Posted
November 3, 2022
Address
2350 Qume Dr, San Jose, CA, 95131-1812

Description

BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BD FACSCanto II, and BD FACSCanto flow cytometers to determine the percentages and absolute counts of human lymphocyte subsets in peripheral whole blood for immunophenotyping.

Reason

Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant.

Action

On 10/03/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via FedEx tracked letters and emails informing customers, through internal investigations, the labels applied to the BD Trucount tubes are becoming detached due to an adhesive change made at a supplier. Customers are instructed to: 1. Immediately review their inventory for the specific catalog and lot numbers. 2. For any tubes where label detachment is observed, BD recommends that the clinician remove the label from the tube and, using a permanent marker, record the tube lot number as well as any other relevant information directly on the tube. Removal of the label prevents it from interfering with any auto-loading sample preparation and analysis systems, as well as from sticking to other tubes. For questions, contact North American Regional Complaint Center at: Phone: 1-844-8BD-LIFE (1-844-823-5433), Say "Recall" when prompted Mon-Fri 8:00am and 5:00pm CT or Email: [email protected]

Distribution

U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay

Quantity

1,457 tubes (corrected on 11/04/2022)