FDA Recall Terminated

ADVIA Centaur Intact PTH Calibrator (6 pack); Catalog Number: SMN 10492387 For in vitro diagnostic use in calibrating Intact PTH assays on the ADVIA Centaur Systems.

Recall: Z-0173-2014 · Initiated August 16, 2013

Recall

Recall Number
Z-0173-2014
Event Number
66132
Firm
Siemens Healthcare Diagnostics, Inc
FEI Number
1219913
Product Code
CEW
Status
Terminated
Root Cause
Packaging
Initiated
August 16, 2013
Posted
November 6, 2013
Terminated
February 13, 2014
Address
333 Coney St, East Walpole, MA, 02032-1516

Description

ADVIA Centaur Intact PTH Calibrator (6 pack); Catalog Number: SMN 10492387 For in vitro diagnostic use in calibrating Intact PTH assays on the ADVIA Centaur Systems.

Reason

PTH Calibrator packaged with only high calibrator.

Action

Packaging error was brought to Siemens' attention through escalations received from (3) customers in the United States. Internal Investigation confirmed the issue was limited to the five (5) kits distributed to these customers. As the customers were aware of the issue and non affected product was provided to the customers to use during Siemens' investigation, no further communication was warranted.

Distribution

Nationwide Distribution including MD, PA, and CT.

Quantity

595 kits total (5 kits with packaging error)