FDA Recall Terminated

VITROS VANC (vancoymycin) Reagent, REF #680 1709. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626

Recall: Z-0173-06 · Initiated October 12, 2005

Recall

Recall Number
Z-0173-06
Event Number
33896
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
LEH
Status
Terminated
Root Cause
Other
Initiated
October 12, 2005
Posted
November 18, 2005
Terminated
September 22, 2008
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS VANC (vancoymycin) Reagent, REF #680 1709. Firm on label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626

Reason

Negatively biased results can be observed when the reagent pack has less than five tests remaining and has been on the analyzer greater than 2 days.

Action

Letters dated 10/12/2005 issued to end users in the U.S. Customers advised they may continue to use the VANC Reagent if their laboratory processes QC fluids according to protocol. A return reply form is included. Foreign affiliates were notified via e-mail on 10/12/2005.

Distribution

Product shipped to medical facilities (end-users) in the U.S. and to 9 International J&J affiliates in Australia, Canada, Caribbean, Chile, France, Philippines, Portugal, Taiwan, UK/Northern Europe, Israel, Saudi Arabia and Venezuela.

Quantity

784 boxes x 6 packs x 50 tests