FDA Recall Terminated

AxSYM Analyzer, Small Heater Block Clamp, manufactured by Abbott Laboratories, Irving, Texas

Recall: Z-0173-04 · Initiated June 6, 2002

Recall

Recall Number
Z-0173-04
Event Number
27179
Firm
Abbott Laboratories, Inc
FEI Number
1628664
Product Code
JJE
Status
Terminated
Root Cause
Other
Initiated
June 6, 2002
Posted
July 20, 2004
Terminated
August 9, 2004
Address
1921 Hurd Drive PO Box 152020, Irving, TX, 75015-2020

Description

AxSYM Analyzer, Small Heater Block Clamp, manufactured by Abbott Laboratories, Irving, Texas

Reason

Clamp can cause misaligned seating of heater block resulting in inaccurate dispension of MEIA Wash Buffer or MUP into the reaction cell. Causes inconsistent results.

Action

All consignees received Technical Service Bulletin (TSB) #83-081. This TSB gave instructions on how to install the AxSYM Small Heater Block clamp upgrade implemented June 6, 2002.

Distribution

Nationwide and to the following foreign locations: Mexico, Viet Nam, Venezuela, Columbia, Argentina, Chile, Japan, Austria, Denmark, France, Israel, Norway, Saudi Arabia, Spain, Turkey, India, Phillippines, Brazil, Puerto Rico, Dominicam Republic, Canada, Germany, Taiwan, Thailand, Belgium, Egypt, Greece, Italy, Poland, South Africa, Sweden, United Arab Emirates, Malaysia, Singapore, Hong Kong, Australia, New Zealand, China, Uruguay, Korea, England, Czech Republic, Finland, Hungary, Netherlands, Portugal, Eastern Europe, Switzerland, Central America, and Pakistan

Quantity

15437 units worldwide