AxSYM Analyzer, Small Heater Block Clamp, manufactured by Abbott Laboratories, Irving, Texas
Recall
- Recall Number
- Z-0173-04
- Event Number
- 27179
- Firm
- Abbott Laboratories, Inc
- FEI Number
- 1628664
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 6, 2002
- Posted
- July 20, 2004
- Terminated
- August 9, 2004
- Address
- 1921 Hurd Drive PO Box 152020, Irving, TX, 75015-2020
Description
AxSYM Analyzer, Small Heater Block Clamp, manufactured by Abbott Laboratories, Irving, Texas
Clamp can cause misaligned seating of heater block resulting in inaccurate dispension of MEIA Wash Buffer or MUP into the reaction cell. Causes inconsistent results.
All consignees received Technical Service Bulletin (TSB) #83-081. This TSB gave instructions on how to install the AxSYM Small Heater Block clamp upgrade implemented June 6, 2002.
Nationwide and to the following foreign locations: Mexico, Viet Nam, Venezuela, Columbia, Argentina, Chile, Japan, Austria, Denmark, France, Israel, Norway, Saudi Arabia, Spain, Turkey, India, Phillippines, Brazil, Puerto Rico, Dominicam Republic, Canada, Germany, Taiwan, Thailand, Belgium, Egypt, Greece, Italy, Poland, South Africa, Sweden, United Arab Emirates, Malaysia, Singapore, Hong Kong, Australia, New Zealand, China, Uruguay, Korea, England, Czech Republic, Finland, Hungary, Netherlands, Portugal, Eastern Europe, Switzerland, Central America, and Pakistan
15437 units worldwide