FDA Recall Terminated

AST R1 Reagent, 120 ml., Pointe Scientific, Inc., Canton, MI; Catalog no. 8-A7561-R1-120. For the quantitative determination of aspartate amoinotransferase in human serum

Recall: Z-0172-2009 · Initiated August 25, 2008

Recall

Recall Number
Z-0172-2009
Event Number
49131
Firm
Pointe Scientific, Inc.
FEI Number
1827821
Product Code
CIT
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
August 25, 2008
Posted
October 22, 2008
Terminated
November 27, 2009
Address
5449 Research Dr, Canton, MI, 48188-2261

Description

AST R1 Reagent, 120 ml., Pointe Scientific, Inc., Canton, MI; Catalog no. 8-A7561-R1-120. For the quantitative determination of aspartate amoinotransferase in human serum

Reason

The R1 component may be contaminated with Serratia liquefaciens, resulting decreased absorbances and failure of the reagent to produce test results.

Action

Consignees were notified via recall letter dated 8/22/08 and mailed 8/25/08 to examine their inventory, dispose of any recalled product found, and to notify the firm for replacements. Contact Pointe Scientific Inc., at 1-800-757-5313 for assistance.

Distribution

Nationwide, Bolivia, Ecuador, Greece, Indonesia, Mexico, Nepal, Philippines, Romania and Serbia.

Quantity

25 vials