FDA Recall Terminated

ARCHITECT Processing Module, manufactured by Abbott Laboratories, Irving, Texas

Recall: Z-0172-04 · Initiated May 8, 2002

Recall

Recall Number
Z-0172-04
Event Number
27180
Firm
Abbott Laboratories, Inc
FEI Number
1628664
Product Code
JJE
Status
Terminated
Root Cause
Other
Initiated
May 8, 2002
Posted
July 20, 2004
Terminated
November 26, 2003
Address
1921 Hurd Drive PO Box 152020, Irving, TX, 75015-2020

Description

ARCHITECT Processing Module, manufactured by Abbott Laboratories, Irving, Texas

Reason

If the interlock cable connected to the door lid sensor is disconnected or fails, the lid is detected as closed even though it may be open.

Action

Firm sent technical service bulletins with corrective instructions to consignees with all existing ARCHITECT i 2000 processing modules in February 2003.

Distribution

States of Virginia, Illinois, New York, Texas, Oklahoma, Pennsylvania, ILlinois, Michigan, California, New Hampshire, Ohio, Florida, Maryland, Nevada, Georgia, Minnesota, Colorado, Arizona, Missouri, Louisiana, Oregon, Wisconsin, Hawaii, Connecticut, New Hampshire, Massachusetts, Washington, South Carolina, Mississippi, Alabama, and Tennessee, and the following foreign countries: Austria, Denmark, Germany, Italy, Poland, South Africa, Sweden, England, Colombia, Australia, Japan, New Zealand, Belgium, Finland, Greece, Netherlands, Portugal, Romania, Turkey, Canada, Mexico, Hong Kong, Korea, Taiwan, Czech Republic, France, Israel, Norway, Saudi Arabia, Spain, United Arab Emirates, Brazil, Venezuela, India, Malaysia, and Thailand.

Quantity

1,167 units worldwide