FDA Recall
Terminated
Liquid AST (SGOT) Reagent Set, Manufactured by Pointe Scientific, Inc., Canton, MI in 150 ml (Product no. A7561-150) and 450 ml (Product no A7561-450). For the quantitative determination of aspartate amoinotransferase in human serum
Recall: Z-0171-2009
·
Initiated August 25, 2008
Recall
- Recall Number
- Z-0171-2009
- Event Number
- 49131
- Firm
- Pointe Scientific, Inc.
- FEI Number
- 1827821
- Product Code
- CIT
- Status
- Terminated
- Root Cause
- Material/Component Contamination
- Initiated
- August 25, 2008
- Posted
- October 22, 2008
- Terminated
- November 27, 2009
- Address
- 5449 Research Dr, Canton, MI, 48188-2261
Description
Liquid AST (SGOT) Reagent Set, Manufactured by Pointe Scientific, Inc., Canton, MI in 150 ml (Product no. A7561-150) and 450 ml (Product no A7561-450). For the quantitative determination of aspartate amoinotransferase in human serum
Reason
The R1 component may be contaminated with Serratia liquefaciens, resulting decreased absorbances and failure of the reagent to produce test results.
Action
Consignees were notified via recall letter dated 8/22/08 and mailed 8/25/08 to examine their inventory, dispose of any recalled product found, and to notify the firm for replacements. Contact Pointe Scientific Inc., at 1-800-757-5313 for assistance.
Distribution
Nationwide, Bolivia, Ecuador, Greece, Indonesia, Mexico, Nepal, Philippines, Romania and Serbia.
Quantity
109 kits