FDA Recall Terminated

Safe Spot Infant Resuscitator, SS3200 Series, Single Patient Use, Manufactured by: Ventlab Corporation, 155 Boyce Drive, Mockville, NC 27028. The Disposable Resuscitators with optional built-in pressure monitor is intended to provide respiratory support in the presence of reversible apnea, commonly associated with cardiopulmonary arrest.

Recall: Z-0170-2013 · Initiated July 11, 2012

Recall

Recall Number
Z-0170-2013
Event Number
63347
Firm
Ventlab Corporation
FEI Number
1314417
Product Code
BTM
Status
Terminated
Root Cause
Component design/selection
Initiated
July 11, 2012
Posted
November 1, 2012
Terminated
June 30, 2013
Address
155 Boyce Dr, Mocksville, NC, 27028-4187

Description

Safe Spot Infant Resuscitator, SS3200 Series, Single Patient Use, Manufactured by: Ventlab Corporation, 155 Boyce Drive, Mockville, NC 27028. The Disposable Resuscitators with optional built-in pressure monitor is intended to provide respiratory support in the presence of reversible apnea, commonly associated with cardiopulmonary arrest.

Reason

Possible volume leakage through the inlet valve during compressions of the resuscitator bag.

Action

VentLab Corporation sent an Attention: Recall Notification letter dated July 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for the affected product, complete and return the enclosed response form by fax to 336-753-5002, indicating whether or not their facility has any of the affected product. Customers can also scan and email the response form to [email protected]. For questions regarding this recall call 336-753-5000.

Distribution

Nationwide Distribution including IL, OH, NY and TN.

Quantity

14,602 total for all units