McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-084-000, Sterile EO, Rx Only
Recall
- Recall Number
- Z-0169-2024
- Event Number
- 92895
- Firm
- Covidien
- FEI Number
- 2936999
- Product Code
- CCW
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- August 10, 2023
- Posted
- November 1, 2023
- Address
- 6135 Gunbarrel Ave, Boulder, CO, 80301-3214
Description
McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-084-000, Sterile EO, Rx Only
Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
On August 10, 2023 Medtronic ( Parent company of Covidien) issued an Urgent via UPS. Medtronic ask consignees to take the following actions: 1. Quarantine all unused product from the affected lots of McGRATH MAC 2 disposable laryngoscope blades. See attachment A for guidance on identifying potentially affected devices. 2. Return all unused product from the affected lots in your inventory to Medtronic as described on the Customer Confirmation Form. 3. Please complete and return the enclosed Customer Confirmation Form even if you do not have unused inventory. 4. Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product from the specified lots has been transferred or distributed. 5. Adverse reactions or quality problems experienced with this product should be reported to FDA and Medtronic: . Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or . Call FDA (800) FDA-1088 . Call Medtronic Quality Assurance at 800-255-6774 option 1, then option 1
US: AK AL AR AZ CA CO CT DE FL GA IA IL IN KS LA MA MD MI MO MT NC ND NH NJ NM NY OH OK OR PA SD TN TX UT VA WA WI WY OUS: Australia Austria Azerbaijan Bahamas Belgium Croatia Czech Republic Denmark Finland France French Guiana Germany Guadeloupe Hong Kong Hungary Ireland Israel Italy Kazakhstan Korea, Republic Of Malta Mayotte Netherlands New Zealand Norway Pakistan Poland Portugal Reunion Romania Spain Sweden Switzerland Turkmenistan United Arab Emirates
42,367