FDA Recall Terminated

Byrd Screw, 2.0X6MM SD, Model Number: 01-0264 Product labeling reads in part as follows: REF: 01-0264, QTY,UOM 00001, BYRD SCREW, 2 OX6MM SELF-DRILLING, 12CM, (4-3/4) (L), STAINLESS STEEL, LOT 999999, (01) 00841036007157(10)999999(13)090201(30)13, 2009-02, ***NON-STERILE PRODUCT***DD 14 03 1, BIOMET MICROFIXATION, INC., 150 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA Instrument used to pull up the broken bone, so it is flush and can be plated fixated. The Byrd screw is then removed.

Recall: Z-0169-2010 · Initiated February 19, 2009

Recall

Recall Number
Z-0169-2010
Event Number
52639
Firm
Biomet Microfixation, Inc.
FEI Number
1032347
Product Code
LXH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 19, 2009
Posted
November 10, 2009
Terminated
May 3, 2012
Address
1520 Tradeport Dr, Jacksonville, FL, 32218-2480

Description

Byrd Screw, 2.0X6MM SD, Model Number: 01-0264 Product labeling reads in part as follows: REF: 01-0264, QTY,UOM 00001, BYRD SCREW, 2 OX6MM SELF-DRILLING, 12CM, (4-3/4) (L), STAINLESS STEEL, LOT 999999, (01) 00841036007157(10)999999(13)090201(30)13, 2009-02, ***NON-STERILE PRODUCT***DD 14 03 1, BIOMET MICROFIXATION, INC., 150 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA Instrument used to pull up the broken bone, so it is flush and can be plated fixated. The Byrd screw is then removed.

Reason

Byrd Screw, Self Drilling, may break during usage. BIOMET has determined that the parts were not correctly processed by the vendor , i.e. that the parts were not hardened in the final processing and that lead to weaker than acceptable material.

Action

Biomet plans to send letters to customers on February 19, 2009. Customers are told to discontinue use of the item and return it to Biomet Microfixation. The firm plans to follow up with an e-mail to their customers. Questions pertaining to this issue, should be directed to Christy Cain at 1.800.874.7711 or 904.741.4400.

Distribution

AZ, CA, GA, IL, IN, NM, NY, TX, VA, Puerto Rico and Canada.

Quantity

22