Volcano s5/s5i imaging system Family Software version 3.2, Part Numbers 435-0602.01 to 05, including ChromaFlo and Virtual Histology, manufactured by Volcano Corporation of Rancho Cordova, CA.
Recall
- Recall Number
- Z-0168-2011
- Event Number
- 56794
- Firm
- Volcano Corporation
- FEI Number
- 2939520
- Product Code
- IYO
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- June 29, 2010
- Posted
- November 1, 2010
- Terminated
- December 22, 2010
- Address
- 2870 Kilgore Rd, Rancho Cordova, CA, 95670-6133
Description
Volcano s5/s5i imaging system Family Software version 3.2, Part Numbers 435-0602.01 to 05, including ChromaFlo and Virtual Histology, manufactured by Volcano Corporation of Rancho Cordova, CA.
Software defect results in improper image brightness, which may result in introduction of artifacts, which may cause misclassification of results and affect classification of tissue types. This may result in misdiagnosis and inappropriate treatment.
Volcano issued a Voluntary Recall of Volcano s5 Family Software version 3.2 letter dated June 30, 2010, identifying the software defect and actions to be taken by Volcano. Field Service Engineers will visit each site and reinstall the previous software version, which will resolve the issue. Customers can contact Volcano at 916 281-2790.
Worldwide Distribution: USA, and the countries of France, Germany, Japan, and the UK.
79 software kits