FDA Recall Terminated

Volcano s5/s5i imaging system Family Software version 3.2, Part Numbers 435-0602.01 to 05, including ChromaFlo and Virtual Histology, manufactured by Volcano Corporation of Rancho Cordova, CA.

Recall: Z-0168-2011 · Initiated June 29, 2010

Recall

Recall Number
Z-0168-2011
Event Number
56794
Firm
Volcano Corporation
FEI Number
2939520
Product Code
IYO
Status
Terminated
Root Cause
Software design
Initiated
June 29, 2010
Posted
November 1, 2010
Terminated
December 22, 2010
Address
2870 Kilgore Rd, Rancho Cordova, CA, 95670-6133

Description

Volcano s5/s5i imaging system Family Software version 3.2, Part Numbers 435-0602.01 to 05, including ChromaFlo and Virtual Histology, manufactured by Volcano Corporation of Rancho Cordova, CA.

Reason

Software defect results in improper image brightness, which may result in introduction of artifacts, which may cause misclassification of results and affect classification of tissue types. This may result in misdiagnosis and inappropriate treatment.

Action

Volcano issued a Voluntary Recall of Volcano s5 Family Software version 3.2 letter dated June 30, 2010, identifying the software defect and actions to be taken by Volcano. Field Service Engineers will visit each site and reinstall the previous software version, which will resolve the issue. Customers can contact Volcano at 916 281-2790.

Distribution

Worldwide Distribution: USA, and the countries of France, Germany, Japan, and the UK.

Quantity

79 software kits