FDA Recall Terminated

Captus 3000 Thyroid Uptake System with Portable Stand option. Capintec item number 5430-2130, serial numbers: 91001, 91003 through 91010. Manufactured by Capintec, Inc, Pittsburgh, PA.

Recall: Z-0167-2011 · Initiated September 20, 2010

Recall

Recall Number
Z-0167-2011
Event Number
56975
Firm
Capintec Inc
FEI Number
2518443
Product Code
KPT
Status
Terminated
Root Cause
Employee error
Initiated
September 20, 2010
Posted
October 28, 2010
Terminated
July 5, 2012
Address
620 Alpha Dr, Pittsburgh, PA, 15238

Description

Captus 3000 Thyroid Uptake System with Portable Stand option. Capintec item number 5430-2130, serial numbers: 91001, 91003 through 91010. Manufactured by Capintec, Inc, Pittsburgh, PA.

Reason

The spring arm on portable stand may have been over tightened during assembly, which may result in failure. If the spring arm fails, it will not balance the collimator weight, and the arm will drop to the lowest position.

Action

Capintec issued an Alert letter to all customers identifying the affected device, informing them of the problem and what actions are to be taken by customers. Capintec will send a replacement arm for the system, which is to be installed by the customer using included instructions. The firm included a correction acknowledgement form to be completed and returned by the customer. Customers can contact Capintec concerning this recall at 201 825-9500.

Distribution

Worldwide Distribution: USA in the states of NC and TN, and in the countries of Pakistan, Paraguay, Portugal, Saudi Arabia, and Spain.

Quantity

9 units