FDA Recall Open, Classified

Cardiac Workstation 7000; Model Number: 860441;

Recall: Z-0166-2026 · Initiated September 12, 2025

Recall

Recall Number
Z-0166-2026
Event Number
97612
Firm
Philips North America
FEI Number
3006648320
Product Code
DPS
Status
Open, Classified
Root Cause
Process control
Initiated
September 12, 2025
Posted
October 16, 2025
Address
222 Jacobs St, Cambridge, MA, 02141-2296

Description

Cardiac Workstation 7000; Model Number: 860441;

Reason

Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.

Action

On September 12, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Share this communication with all clinical staff to review and undedrstand. Please this letter with the device. Pass the notice along to all those who need to be aware within the organization or to any organization where affected devices were transferred. Ensure technical solution established by Philips is implemented as soon as available within the timeframe communicated by Philips. In case any device cannot be corrected, inform Philips of the reason why the Technical Solution could not be implemented. Action By Philips: A Philips representative will contact customers to arrange a replacement of the device's hardware on affected devices with the pinched power module wire within the housing of the Cardiac Workstation.

Distribution

US Nationwide distribution in the states of CA, IA, TN, TX, VA.

Quantity

29 units