FDA Recall Terminated

Perkin Elmer Specimen Gate, Screening Center. Data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.

Recall: Z-0166-2016 · Initiated August 24, 2015

Recall

Recall Number
Z-0166-2016
Event Number
72013
Firm
Perkin Elmer Life Sciences Inc Laitun 4 Mustionkatu 6 Turku Finland
FEI Number
3002808498
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
August 24, 2015
Terminated
January 21, 2021

Description

Perkin Elmer Specimen Gate, Screening Center. Data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.

Reason

It would be possible for an abnormal analytical test result value from an affected newborn to be reported with a default result interpretation of normal. The analytical test result value would be correct but the interpretation of the test result value would be incorrect.

Action

Perkin Elmer issued the Recall Letter on 8/24/15 delivered to the customer by email .Software Services will be asked to make sure the customers receive the Recall Letter and implement the recommended corrective measures: Ensure all laboratory routine sample testing procedures are followed to avoid procedural anomalies. In the event a procedure anomaly occurs, review the final analytical result and the corresponding interpretation to confirm both are accurate before releasing the report. Once a PerkinElmer reporting error query is available, continue to run the query until the final corrective action is deployed to your site. The final corrective action will involve an installation update to the software. For further information, please contact your local PerkinElmer representative or [email protected]

Distribution

US distribution to FL, GA, and NV; and Canada

Quantity

6 units