Siemens brand 550 TxT Patient Table, Part No. 7346534, Distributed by and/or Manufactured by Siemens AG, Medical Solutions, Germany
Recall
- Recall Number
- Z-0166-2011
- Event Number
- 56352
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2937457
- Product Code
- JAI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 31, 2010
- Posted
- October 28, 2010
- Terminated
- March 1, 2012
- Address
- 4040 Nelson Ave, Concord, CA, 94520-1200
Description
Siemens brand 550 TxT Patient Table, Part No. 7346534, Distributed by and/or Manufactured by Siemens AG, Medical Solutions, Germany
An increase in voltage may cause a short circuit of a power transistor that may cause a smoldering and smoking effect of the electrical wiring and also an unexpected lowering of the Table. Both issues present the potential for injury to a patient or the operator.
Siemens sent a Customer Information Letter on March 31, 2010, to all consignees. The letter identified the product, the problem, and the action that should be taken. Siemens implemented minor modifications to address the issues. No further action was required for the continual operation of the 550TxT Table. For questions regarding this recall call 925-602-8083.
Worldwide Distribution including USA
326 units