Siemens brand 550 TxT Patient Table, Part No. 7346534, Distributed by and/or Manufactured by Siemens AG, Medical Solutions, Germany

Recall: Z-0166-2011 · Initiated March 31, 2010

Recall

Recall Number
Z-0166-2011
Event Number
56352
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2937457
Product Code
JAI
Status
Terminated
Root Cause
Other
Initiated
March 31, 2010
Posted
October 28, 2010
Terminated
March 1, 2012
Address
4040 Nelson Ave, Concord, CA, 94520-1200

Description

Siemens brand 550 TxT Patient Table, Part No. 7346534, Distributed by and/or Manufactured by Siemens AG, Medical Solutions, Germany

Reason

An increase in voltage may cause a short circuit of a power transistor that may cause a smoldering and smoking effect of the electrical wiring and also an unexpected lowering of the Table. Both issues present the potential for injury to a patient or the operator.

Action

Siemens sent a Customer Information Letter on March 31, 2010, to all consignees. The letter identified the product, the problem, and the action that should be taken. Siemens implemented minor modifications to address the issues. No further action was required for the continual operation of the 550TxT Table. For questions regarding this recall call 925-602-8083.

Distribution

Worldwide Distribution including USA

Quantity

326 units