FDA Recall Terminated

Koodinat M/MP Patient Table used with Angiostar Model 5294009. The device is used as an Angiographic X-Ray System.

Recall: Z-0166-2010 · Initiated July 1, 2009

Recall

Recall Number
Z-0166-2010
Event Number
53353
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IZI
Status
Terminated
Root Cause
Labeling False and Misleading
Initiated
July 1, 2009
Posted
November 10, 2009
Terminated
January 12, 2010
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

Koodinat M/MP Patient Table used with Angiostar Model 5294009. The device is used as an Angiographic X-Ray System.

Reason

The table swivel locking mechanism may not completely engage adn could result in uncontrolled patient movement during transfer resulting in a patient fall.

Action

The recalling firm issued a Customer Safety Advisory Notice dated July 1, 2009 to affected customers via Update Instructions AX025/09/S to inform customer of the potential issues and provide instructions to prevent occurrence. The firm also provided an addendum to User Manual. Direct questions about the recall to Siemens Medical Solutions USA, Inc. by calling 1-610-219-6300.

Distribution

Nationwide Distribution.

Quantity

107 units