FDA Recall
Terminated
Allura Xper FD 20/10 x-ray, and fluoroscopy
Recall: Z-0166-2007
·
Initiated October 12, 2006
Recall
- Recall Number
- Z-0166-2007
- Event Number
- 36573
- Firm
- Philips Medical Systems North America Co. Phillips
- FEI Number
- 1000524572
- Product Code
- IZI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 12, 2006
- Posted
- November 14, 2006
- Terminated
- September 24, 2010
- Address
- 22100 Bothell Everett Hwy, Bothell, WA, 98021-8431
Description
Allura Xper FD 20/10 x-ray, and fluoroscopy
Reason
Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
Action
On 10/12/06 the firm issued a letter to their consignees advising of the problem and how to avoid the problem by using specific image formats. If the system should lock up, consignees are to switch the system off and then restart. The firm will be conducting software upgrades to the devices once software is ready (estimated time for FD 10/10 is November 2006 and estimated time for FD 20/10 is January 2007). Later, the firm identified 8 additional units and issued a letter on 12/11/06 to affected customers.
Distribution
Nationwide.
Quantity
19 devices