FDA Recall Terminated

Allura Xper FD 20/10 x-ray, and fluoroscopy

Recall: Z-0166-2007 · Initiated October 12, 2006

Recall

Recall Number
Z-0166-2007
Event Number
36573
Firm
Philips Medical Systems North America Co. Phillips
FEI Number
1000524572
Product Code
IZI
Status
Terminated
Root Cause
Other
Initiated
October 12, 2006
Posted
November 14, 2006
Terminated
September 24, 2010
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

Allura Xper FD 20/10 x-ray, and fluoroscopy

Reason

Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.

Action

On 10/12/06 the firm issued a letter to their consignees advising of the problem and how to avoid the problem by using specific image formats. If the system should lock up, consignees are to switch the system off and then restart. The firm will be conducting software upgrades to the devices once software is ready (estimated time for FD 10/10 is November 2006 and estimated time for FD 20/10 is January 2007). Later, the firm identified 8 additional units and issued a letter on 12/11/06 to affected customers.

Distribution

Nationwide.

Quantity

19 devices