FDA Recall Terminated

Roche/Hitachi Modular E Module immunoassay analyzer, Roche Diagnostics, Ind., IN; GMMI Nos. 04998642001 and 03617505001.

Recall: Z-0165-2009 · Initiated June 24, 2008

Recall

Recall Number
Z-0165-2009
Event Number
48711
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
JJE
Status
Terminated
Root Cause
Other
Initiated
June 24, 2008
Posted
October 22, 2008
Terminated
October 22, 2009
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Roche/Hitachi Modular E Module immunoassay analyzer, Roche Diagnostics, Ind., IN; GMMI Nos. 04998642001 and 03617505001.

Reason

A software bug may result in pipetting from an incorrect reagent pack and/or assigning calibration curve parameters incorrectly.

Action

The consignees were notified of the problem via letter dated 6/24/08 and were provided instructions for a workaround.

Distribution

Nationwide Distribution --- including states of Florida, New Jersey and North Carolina.