FDA Recall Terminated

Transfer/Gait Belt, Plastic, Wipeable, packaged in plastic bag, 36 plastic bags per box.

Recall: Z-0164-2018 · Initiated August 10, 2017

Recall

Recall Number
Z-0164-2018
Event Number
78363
Firm
Medline Industries Inc
FEI Number
1417592
Product Code
IKX
Status
Terminated
Root Cause
Device Design
Initiated
August 10, 2017
Terminated
April 24, 2020
Address
Three Lakes Drive, Northfield, IL, 60093

Description

Transfer/Gait Belt, Plastic, Wipeable, packaged in plastic bag, 36 plastic bags per box.

Reason

The locking mechanism on the belt may fail to secure/hold the strap when in locked position potentially allowing the belt strap to slip through the buckle.

Action

Medline Industries sent an Immediate Action Required Recall Notification letter via First Class mail dated August 10, 2017. The recall was later expanded and customers were notified via letter of the additional lots dated August 21, 2017. Instructions in both letters included to check inventory and quarantine any affected product, complete and return the included response form, coordinate the return of all affected product, and notify customers if the product was further distributed. Additional customers were identified and notified of the recall via the same process on September 19th, 2017. For further questions, please call (866) 359-1704.

Distribution

USA (nationwide) Distribution to the states of : AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NV, NY, OH , OK, OR, PA, RI, SC, TN, TX, VA, VT, WA, WI and WV.

Quantity

1,428 units total