Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Ct., Allendale NJ 07401-1677 USA +1-201-750-8000 Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.
Recall
- Recall Number
- Z-0163-2013
- Event Number
- 63272
- Firm
- Stryker Spine
- FEI Number
- 3004024955
- Product Code
- LXH
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 5, 2012
- Posted
- October 29, 2012
- Terminated
- January 28, 2013
- Address
- 2 Pearl Ct, Allendale, NJ, 07401-1611
Description
Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Ct., Allendale NJ 07401-1677 USA +1-201-750-8000 Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.
Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.
Stryker Spine sent an Urgent Medical Device Field Correction Notification letter beginning July 5, 2012, to all affected customers. Teh letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check the drill bit length in each respective drill guide prior to surgery to ensure the correct length protrudes through the guide. Customers with questions should call 201-760-8298. Customers were asked to report any adverse events or product quality problems to 1-888-457-7463.
Nationwide distribution.
6154 units