FDA Recall Terminated

Signature Patient-Specific Surgical Guides; patient specific instruments used as a guide during surgical procedure for total and partial knee arthroplasty. Part of the Signature Personalized Patient Care System.

Recall: Z-0162-2016 · Initiated September 23, 2015

Recall

Recall Number
Z-0162-2016
Event Number
72425
Firm
Materialise N.V. 15 Technologielaan Heverlee Belgium
FEI Number
3003998208
Product Code
OSF
Status
Terminated
Root Cause
Use error
Initiated
September 23, 2015
Posted
October 23, 2015
Terminated
February 12, 2016

Description

Signature Patient-Specific Surgical Guides; patient specific instruments used as a guide during surgical procedure for total and partial knee arthroplasty. Part of the Signature Personalized Patient Care System.

Reason

Potential inaccuracies in the alignment of the guides. Surgery was inaccurately planned because the patient image data was incorrectly processed.

Action

On September 16, 2015 Materialise NV identified four cases where not all available image sets were used to design the case. On September 18, 2015, prior to the surgery dates Materialise informed (phone call) the Signature Services Manager of Zimmer Biomet that inaccuracies in the alignment of the knee might be observed with the use of the guides due to this issue. Biomet was asked to segregate the packages for these four cases and to return them to the Materialise US office. On September 23, 2015 this was followed up by e-mail notification letter with response forms. Biomet was requested to contact Materialise customer service to return these four cases and arrange a replacement set of patient specific surgical guides Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time. E-mail : ortho-quality@materialise. be Customer Service Materialise USA LLC 44650 Helm Court Plymouth, Michigan 48170 734-259-7033

Distribution

Distributed in MI, ID, TX, and in France.

Quantity

4