FDA Recall
Terminated
''Fine Tuning Your Pump'' booklet provided with Model MMT-515 AND MMT-715 INSULIN INFUSION PUMPS
Recall: Z-0162-06
·
Initiated April 27, 2005
Recall
- Recall Number
- Z-0162-06
- Event Number
- 33025
- Firm
- Medtronic MiniMed
- FEI Number
- 3003166194
- Product Code
- LZG
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 27, 2005
- Posted
- November 18, 2005
- Terminated
- March 20, 2007
- Address
- 18000 Devonshire St, Northridge, CA, 91325-1219
Description
''Fine Tuning Your Pump'' booklet provided with Model MMT-515 AND MMT-715 INSULIN INFUSION PUMPS
Reason
The text description of how to perform insulin sensitivity testing and associated worksheet included in the booklet that is shipped with the Paradigm 515 and Paradigm 715 insulin pumps contained a basic flaw in that the blood glucose after the correction should be compared to the target glucose and it says compare to the beginning blood glucose in the procedure.
Action
All end users who recieved the model MMT-515 or MMT-715 pump that included the 'Fine Yuning Your Pump' workbook will be notified through a letter.
Distribution
Nationwide, Bahamas, Bermuda, Canada, Mexico, and Puerto Rico.
Quantity
Approximately 14,560