FDA Recall Terminated

''Fine Tuning Your Pump'' booklet provided with Model MMT-515 AND MMT-715 INSULIN INFUSION PUMPS

Recall: Z-0162-06 · Initiated April 27, 2005

Recall

Recall Number
Z-0162-06
Event Number
33025
Firm
Medtronic MiniMed
FEI Number
3003166194
Product Code
LZG
Status
Terminated
Root Cause
Other
Initiated
April 27, 2005
Posted
November 18, 2005
Terminated
March 20, 2007
Address
18000 Devonshire St, Northridge, CA, 91325-1219

Description

''Fine Tuning Your Pump'' booklet provided with Model MMT-515 AND MMT-715 INSULIN INFUSION PUMPS

Reason

The text description of how to perform insulin sensitivity testing and associated worksheet included in the booklet that is shipped with the Paradigm 515 and Paradigm 715 insulin pumps contained a basic flaw in that the blood glucose after the correction should be compared to the target glucose and it says compare to the beginning blood glucose in the procedure.

Action

All end users who recieved the model MMT-515 or MMT-715 pump that included the 'Fine Yuning Your Pump' workbook will be notified through a letter.

Distribution

Nationwide, Bahamas, Bermuda, Canada, Mexico, and Puerto Rico.

Quantity

Approximately 14,560