FDA Recall Terminated

Accessory Head Rest for Alphastar Operating Tables

Recall: Z-0161-2009 · Initiated September 26, 2008

Recall

Recall Number
Z-0161-2009
Event Number
49545
Firm
MAQUET Inc.
FEI Number
3008355164
Product Code
FQO
Status
Terminated
Root Cause
Other
Initiated
September 26, 2008
Posted
October 16, 2008
Terminated
April 16, 2009
Address
1140 Us Highway 22, Ste 202, Bridgewater, NJ, 08807-2958

Description

Accessory Head Rest for Alphastar Operating Tables

Reason

During installation or cleaning there is a risk of crushing the finger by handling the head rest in the area of the gas strut while simultaneously actuating the release lever.

Action

Urgent Device Correction letters were sent beginning September 26, 2008 via Fed Ex. Letters identified the potential for finger crush in conjunction with the head rest. Proper operating instructions are provided as well as a warning sticker to be applied to the head rest indicating proper positioning. All letters included a customer response form to be completed by each institution. Information requested is: that device correction letter has been read and instructions followed; that the corrected operating instructions have been received; that the warning sticker has been properly placed on the headrest, or that they do not have the head rest identified in the device correction letter. Questions should be addressed to USA Maquet Inc at 1-888-627-8383 #1.

Distribution

Nationwide Distribution

Quantity

88 units in US; 1884 units worldwide