Accessory Head Rest for Alphastar Operating Tables
Recall
- Recall Number
- Z-0161-2009
- Event Number
- 49545
- Firm
- MAQUET Inc.
- FEI Number
- 3008355164
- Product Code
- FQO
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 26, 2008
- Posted
- October 16, 2008
- Terminated
- April 16, 2009
- Address
- 1140 Us Highway 22, Ste 202, Bridgewater, NJ, 08807-2958
Description
Accessory Head Rest for Alphastar Operating Tables
During installation or cleaning there is a risk of crushing the finger by handling the head rest in the area of the gas strut while simultaneously actuating the release lever.
Urgent Device Correction letters were sent beginning September 26, 2008 via Fed Ex. Letters identified the potential for finger crush in conjunction with the head rest. Proper operating instructions are provided as well as a warning sticker to be applied to the head rest indicating proper positioning. All letters included a customer response form to be completed by each institution. Information requested is: that device correction letter has been read and instructions followed; that the corrected operating instructions have been received; that the warning sticker has been properly placed on the headrest, or that they do not have the head rest identified in the device correction letter. Questions should be addressed to USA Maquet Inc at 1-888-627-8383 #1.
Nationwide Distribution
88 units in US; 1884 units worldwide