FDA Recall Terminated

ICCG AC-Powered Adjustable Medical Bed

Recall: Z-0158-06 · Initiated July 23, 2004

Recall

Recall Number
Z-0158-06
Event Number
29665
Firm
Invacare Corporation
FEI Number
3002889431
Product Code
FNL
Status
Terminated
Root Cause
Other
Initiated
July 23, 2004
Posted
November 17, 2005
Terminated
April 12, 2012
Address
1 Invacare Way, Elyria, OH, 44035

Description

ICCG AC-Powered Adjustable Medical Bed

Reason

Component Defect. Metal Oxide Varister (MOV) component of the junction box may short circuit with the potential for device failure.

Action

The manufacturer notified their distributors of the recall with telephone calls & a letter. The distributors were provided information about the recall and informed of their recall responsibilites.

Distribution

World wide

Quantity

14,343 units