Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085

Recall: Z-0156-2024 · Initiated February 28, 2023

Recall

Recall Number
Z-0156-2024
Event Number
93018
Firm
Omega Medical Imaging, Inc.
FEI Number
1000222289
Product Code
IWE
Status
Open, Classified
Root Cause
Device Design
Initiated
February 28, 2023
Posted
October 24, 2023
Address
3400 Saint Johns Pkwy, Ste 1020, Sanford, FL, 32771-6769

Description

Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085

Reason

An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.

Action

Omega Medical Imaging issued a Field Safety Advisory Notice to its consignees on 02/28/2023 by email. The notice explained the issue and pledged to evaluate all potentially impacted devices and mitigate the issue to bring the boom systems up to the latest safety design and standard of the newer models approved for field use.

Distribution

Worldwide distribution - US Nationwide and the country of Pakistan.

Quantity

52 units