FDA Recall Terminated

Precision Medical- Easy Mate Portable Liquid Oxygen System (container), PM22010 Series, Model number 2201, Precision Medical, Inc., Northampton, PA 18067

Recall: Z-0153-2007 · Initiated October 4, 2006

Recall

Recall Number
Z-0153-2007
Event Number
36547
Firm
Precision Medical, Inc.
FEI Number
2523148
Product Code
BYJ
Status
Terminated
Root Cause
Other
Initiated
October 4, 2006
Posted
November 4, 2006
Terminated
February 15, 2008
Address
300 Held Dr, Northampton, PA, 18067-1150

Description

Precision Medical- Easy Mate Portable Liquid Oxygen System (container), PM22010 Series, Model number 2201, Precision Medical, Inc., Northampton, PA 18067

Reason

Component assembly; The fill chuck is not assembled properly which may allow the device to come apart over time releasing its contents.

Action

The recalling firm telephoned the company to inform them of the problem and to make them aware that replacements would be made available starting on 10/16/06.

Distribution

The product was shipped to home care providers nationwide

Quantity

550 units