FDA Recall Terminated

Neos Vanadium Excavator Part Number 591/1

Recall: Z-0152-2010 · Initiated November 14, 2008

Recall

Recall Number
Z-0152-2010
Event Number
50763
Firm
Sybron Dental Specialties
FEI Number
2024312
Product Code
EKC
Status
Terminated
Root Cause
Component change control
Initiated
November 14, 2008
Posted
November 9, 2009
Terminated
August 9, 2012
Address
1717 W Collins Ave, Orange, CA, 92867-5422

Description

Neos Vanadium Excavator Part Number 591/1

Reason

The tips of the Neos Vanadium Excavators may prematurely release from the handle during use, and if so, the tip may be accidentally swallowed.

Action

Sybron Dental wrote to the distributor by letter dated November 20, 2008 and posted Nov 24, 2008 titled "URGENT MEDICAL DEVICE RECALL," informing them of the recall, and the reason for recall. Sybron informed the distributor to return affected batches of product. Any product in stock with the mentioned batch number was replaced at no charge or credited to their account. The consignee was also asked to contact KerrHawe Customer Care at +41 91 610 05 05 to receive an RGA number. The affected product should be returned to KerrHawe at the following address: KerrHawe SA Via Strecce 4 CH-6934 Bioggio Switzerland PLEASE LABEL RETURNED PRODUCT "RECALLED PRODUCT RETURN". In addition, they asked for the completed RETURN FORM to be returned by fax to +41 91 610 05 14

Distribution

Worldwide distribution - Austria, Belgium, Denmark, France, Germany, Italy, Netherlands, Portugal, Spain, Switzerland, China, and the United States. One US consignee.

Quantity

72