FDA Recall Terminated

Hitachi Echelon Magnetic Resonance Imaging System, Serial Number: V-001, equipped with CTL coil type: MR-CTL-150.

Recall: Z-0152-2007 · Initiated August 14, 2006

Recall

Recall Number
Z-0152-2007
Event Number
36525
Firm
Hitachi Medical Systems America Inc
FEI Number
1528028
Product Code
MOS
Status
Terminated
Root Cause
Other
Initiated
August 14, 2006
Posted
November 4, 2006
Terminated
September 19, 2012
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087-2371

Description

Hitachi Echelon Magnetic Resonance Imaging System, Serial Number: V-001, equipped with CTL coil type: MR-CTL-150.

Reason

Component defect/overheating device; A malfunctioning decoupling circuit in the CTL coil caused high current flow and circuit overheating.

Action

The recalling firm issued a Device Correction letter to the clinic customer on 9/15/2006 making them aware of this potential problem and of their completed corrective action on 8/14/2006, by a Service Rep visit.

Distribution

The recalled unit was distributed to a single customer.

Quantity

1 unit