Hitachi Echelon Magnetic Resonance Imaging System, Serial Number: V-001, equipped with CTL coil type: MR-CTL-150.
Recall: Z-0152-2007
·
Initiated August 14, 2006
Recall
- Recall Number
- Z-0152-2007
- Event Number
- 36525
- Firm
- Hitachi Medical Systems America Inc
- FEI Number
- 1528028
- Product Code
- MOS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 14, 2006
- Posted
- November 4, 2006
- Terminated
- September 19, 2012
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087-2371
Description
Hitachi Echelon Magnetic Resonance Imaging System, Serial Number: V-001, equipped with CTL coil type: MR-CTL-150.
Reason
Component defect/overheating device; A malfunctioning decoupling circuit in the CTL coil caused high current flow and circuit overheating.
Action
The recalling firm issued a Device Correction letter to the clinic customer on 9/15/2006 making them aware of this potential problem and of their completed corrective action on 8/14/2006, by a Service Rep visit.
Distribution
The recalled unit was distributed to a single customer.
Quantity
1 unit