FDA Recall Terminated

HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients

Recall: Z-0151-2016 · Initiated April 22, 2014

Recall

Recall Number
Z-0151-2016
Event Number
72319
Firm
Hamilton Medical, Inc.
FEI Number
2937708
Product Code
CBK
Status
Terminated
Root Cause
Device Design
Initiated
April 22, 2014
Posted
November 4, 2015
Terminated
August 24, 2017
Address
4990 Energy Way, Reno, NV, 89502-4123

Description

HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients

Reason

Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group (neonatal, pediatric and adult).

Action

Hamilton Medical issued a letter dated April 22, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that their Hamilton Medical Account Manager would contact them to coordinate the deactivation of the Ventilation Suppression feature. Customers were asked to review the Medical Device Field Safety Corrective Action. Customers with questions were instructed to call 800-426-6331, Ext 215.

Distribution

Nationwide Distribution

Quantity

1128