Joerns EasyCare Bed Product Usage: The AC-powered adjustable hospital bed is a device intended for medical purposes. The bed includes a built-in electric motor and remote controls that can be operated by the patient to adjust the height and surface contour of the bed.
Recall
- Recall Number
- Z-0151-2015
- Event Number
- 69400
- Firm
- Joerns Healthcare
- FEI Number
- 3009402404
- Product Code
- FNL
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- September 24, 2014
- Posted
- October 30, 2014
- Terminated
- February 1, 2016
- Address
- 2100 Design Rd, Arlington, TX, 76014-4593
Description
Joerns EasyCare Bed Product Usage: The AC-powered adjustable hospital bed is a device intended for medical purposes. The bed includes a built-in electric motor and remote controls that can be operated by the patient to adjust the height and surface contour of the bed.
The Easy-Care Care-Lock feature installed at the foot of the bed may potentially fail to stay engaged (locked) after the Care-Lock System has been engaged by stepping on the red (locked) pedal.
Joerns sent an Urgent: Medical Device Correction letter dated September 24, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. The firm will provide the customers with a temporary solution kit to safegard the bed until a replacement Care-Lock assembly can be shipped and installed. Customers were instructed to complete the response form. For questions call 800-826-0270.