FDA Recall Terminated

VITROS Chemistry Products GLU DT Slides GEN 56, REF (Catalog #) 153 2316, 25 slides per box. VITROS Chemistry Products GLU DT Slides quantitatively measure glucose (GLU) concentration in serum and plasma. Ortho-Clinical Diagnostics, a Johnson & Johnson company. Ortho-Clinical Diagnostics, Inc. Rochester, NY 14626. .

Recall: Z-0151-2007 · Initiated September 27, 2006

Recall

Recall Number
Z-0151-2007
Event Number
36509
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
CGA
Status
Terminated
Root Cause
Other
Initiated
September 27, 2006
Posted
November 4, 2006
Terminated
August 3, 2012
Address
100 Indigo Creek Drive, Rochester, NY, 14626-5101

Description

VITROS Chemistry Products GLU DT Slides GEN 56, REF (Catalog #) 153 2316, 25 slides per box. VITROS Chemistry Products GLU DT Slides quantitatively measure glucose (GLU) concentration in serum and plasma. Ortho-Clinical Diagnostics, a Johnson & Johnson company. Ortho-Clinical Diagnostics, Inc. Rochester, NY 14626. .

Reason

Positively biased results when using the VITROS Chemistry Products GLU DT Slides GEN 56.

Action

On 9/26/06, the recalling firm notified its foreign affiliate consignees of this 'field correction', and instructed them to inform their customers who received these slides of this issue, provided them with an updated Calibration Data Module and instructed to re-calibrate these slides for continued use. On 9/27/06, all US end-use consignees were informed of this issue via Federal Express overnight letter (dated 9/27/06), and instructing them to re-calibrate these slides for continued use after they have installed the Calibration Data Module. The new CDMs were sent to all customers with VITROS DT 60/DT60 II Chemistry systems starting 9/27/06 concurrent with, but separate, from the correction letters. This includes consignees who purchased their GLU DT slides from distributors. On 9/28/06, all distributors were notified of this 'field correction', providing instructions and 'Questions & Answers' for them, and instructing them to inform their customers who received these slides of this issue. Calibration Data Modules were sent to all customers with VITROS DT 60/DT60 II Chemistry Systems.

Distribution

Worldwide Distribution

Quantity

93,252 boxes (25 slides per box) in total