Adjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids. Product Usage: Intended for use in surgical treatment of morbid obesity
Recall
- Recall Number
- Z-0150-2014
- Event Number
- 66465
- Firm
- Ethicon Endo-Surgery Inc
- FEI Number
- 1527736
- Product Code
- LTI
- Status
- Terminated
- Root Cause
- Unknown/Undetermined by firm
- Initiated
- September 30, 2013
- Posted
- November 5, 2013
- Terminated
- March 26, 2014
- Address
- 4545 Creek Rd, Cincinnati, OH, 45242-2803
Description
Adjustable Gastric Band, packaged alone in transparent PETG inner and outer blisters with Tyvek lids. Product Usage: Intended for use in surgical treatment of morbid obesity
During aging studies, the firm determined that the gastric belts were unable to pass the test requirements established for Force to Lock.
Ethicon sent an Urgent: Medical Device Recall letter dated September 30, 2013 to affected consignees via overnight, UPS to allow tracking of the receipt. The letter identified affected products, problem and actions to be taken. Each consignee is requested to respond to a provided Business Reply Form (BRF), indicating whether they have affected product, how much and whether they will/have returned the product or whether they no longer have any of the recall product. For questions contact your local sales representative or call our Customer Support Center 1-877-384-4266 option 6.
Worldwide Distribution: USA (nationwide) including the states of: Alabama, Connecticut, Indiana, Michigan, North Carolina, South Carolina, Alaska, Delaware, Kansas, Minnesota, Ohio, Tennessee, Arizona, Florida, Kentucky, Mississipp,i Oklahoma, Texas, Arkansas, Georgia, Louisiana, Nebraska, Oregon, Virginia, California, Idaho, Maryland, Nevada, Pennsylvania, Washington, Colorado, Illinois, Massachusetts, New Jersey, Rhode Island, and West Virginia; and the countries of: Argentina, France, Malaysia, Spain, Australia, Georgia, Martinique, Sweden, Belgium, Germany, Mexico, Switzerland, Brazil, Greece, Netherland, Taiwan, Chile, Hungary, Panama, Tunisia, China, India, Philippines, United Arab Emirates, Czech Republic, Israel, Poland, United Kingdom, Egypt, Italy, Portugal, Venezuela, Estonia, Kenya, Russia, Finland, Kuwait, and Saudi Arabia.
Product Code: 2200-X - 2,475 units